| SERIES
600 EasyCal¢â Digital
Temperature Loggers;
615 EasyCal¢â
Single Input Thermometer (J,K,E,T) |
|
- CE-Certificated.
- User-friendly front panel is easy to set up and operate.
- Measures J, K, T, and E-types of thermocouples.
- Selectable units of °C, °F, or K (Kelvin).
- Data Logging up to 9,999 points of data with user adjustable
recording interval.
- Recall function allows logged data to be easily reviewed on the meter display.
- Records MAX/MIN/AVG readings for later recall.
- Data Hold function freezes reading on display.
- Auto-Power-Off increases battery life ; typical 300 hrs. with 2 x 1.5 AAA Alkaline battery.
- Real Time Clock & Internal Temperature display.
- The thermocouple(s) with wire spool(s) can be stored in the back of the thermometer.
- Battery door allows easy battery replacement without breaking the calibration seal.
- Transfer measuring or recorded data to a PC via USB serial port for analysis and graphing.
- Windows® compatible software CD and USB Cable.
- 3 year warranty.
- Immersion(30min.) & Dust proof : IP67 according to IEC 60529.
|
|
| SPECIFICATIONS
|
|
Measurement Range
|
J-type : -210 °C to +1200 °C (-346 °F to +2192 °F)
K-type : -200 °C to +1370 °C (-328 °F to +2498 °F)
T-type : -250 °C to +400 °C (-418 °F to +752 °F)
E-type : -150 °C to +1000 °C (-238 °F to +1832 °F)
|
|
Display Resolution
|
J-type : 0.1 °C / K < 550 ° 0.1 °F < 1000 °
1.0 °C / K ¡Ã550 ° 1.0 °F ¡Ã1000 °
K-type : 0.1 °C / K < 750 ° 0.1 °F < 1000 °
1.0 °C / K¡Ã750 ° 1.0 °F¡Ã1000 °
T-type : 0.1 °C / K / °F °
E-type : 0.1 °C / K < 400 ° 0.1 °F < 750 °
1.0 °C / K ¡Ã400 ° 1.0 °F ¡Ã750 °
|
|
Measurement Accuracy
|
J, K, T, and E-type: ± [0.1 % + 0.5 °C (1.0 °F)]
[ Below -100 °C (-148 °F): add 0.2 % of reading for J, K, and E; and 0.5 % of reading for T-type ]
|
|
Temperature Coefficient
|
0.01 % of reading +0.05 °C per °C (0.1 °F per °F) for < +18 °C (+64 °F) or > +28 °C (+82 °F)
[ Below -100 °C (-148 °F): add 0.05 % of reading for J, K, and E; and 0.1 % of reading for T-type ]
|
Real Time Clock Tolerance
|
about 1 sec per day
|
|
Temperature Scale
|
ITS-90 (International Temperature Scale of 1990)
|
|
Applicable Standards
|
N.I.S.T. Monograph 175 revised to ITS-90
|
|
Accuracy is specified for ambient temperatures between 18 °C (64 °F) and 28 °C (82 °F) for a period of 1 year. The above specifications do not include thermocouple error.
|
|
|
|
|
|
| Series 600 Digital Temperature Loggers are most advanced
microprocessor-based, digital thermometers designed to
use external J-, K-, T-, and E-type thermocouples (temperature
probes) as temperature sensors. Series 600 Thermometers
with fast response and laboratory accuracy (0.1 % + 0.5
°C) have logging and PC interface capabilities. The
powerful Windows® compatible software
is optionally provided. Convenient closed case easy calibration
(EasyCal¢â) allows adjustments to be made directly
by using the menu keys. The thermocouple(s) with wire spool(s) can be stored in the back of the thermometer.
The 615 has a single input. |
|
|
|
|
User's Manual download |
|
|
|
|
|
Please
contact us for more information at; |
|
FINE INSTRUMENTS CORP.
Tel : +82-31-470-8888
Fax : +82-31-470-8887
Room # 601, DAEWOO TECHNOPIA, 768-1, WONSI-DONG, DANWON-GU, ANSAN-SI,GYEONGGI-DO, SOUTH KOREA
E-mail : fine@finest.co.kr
|
|